file db/vittara.template
{
pattern
{DEVICE,         mm2Hz, RDBD, KP,   MOTR_PID_SCAN, REF, DIFF, SCAN,     NSAM,  PAD }

{"OSC:LR20:20:", 30150, "1", 0.001,  "10 second",  2,  0.25, "1 second", 100, "LASR:IN20:2:LSR2WTDIFF" }

}

# DEVICE: device prefix
# mm2Hz: conversion factor between motor position and frequency
# RDBD: retry deadband for the coarse motor
# KP: proportional gain for coarse motor feedback
# MOTR_PID_SCAN: motor PID loop update
# REF: Desired piezo position (for coarse PID)
# DIFF: Acceptible piezo difference (for coarse PID)
# SCAN: statistics update
# NSAM: number of samples in the analysis
# PAD: PAD PV