4th July 2011 Version 1_3 Fixed a race condition that occurs when multiple stages are driven by the same ANC350 controller. Mutex protection has been placed around the correlation ID, and the ID is copied to a local variable whilst waiting for the response to allow servicing of requests from other motor records. 10th March 2010 Version 1_2 Some fixes to improve the motor record support were submitted by Emma Shepherd at Diamond Light Source after testing. These fixes are included in this version. 26th October 2009 Version 1_1 This is a patch to solve a problem with mismatching of correlation numbers sent to the ANC350. These numbers are confirmed in the response and a situation would arise where the numbers were larger than could be accepted by the ANC350. The correlation numbers are now wrapped at 10000 to stop this problem from arising.