# Protocol based on MKS Spectra MicroVision IP ASCII Protocol



Terminator = CR LF;
#OutTerminator = CR LR;
#InTerminator= CR CR;
#LockTimeout = 70;
#ReplyTimeout  = 100;
#PollPeriod=$ReplyTimeout;
#ReadTimeout  = 100;
#WriteTimeout = 150;
#ExtraInput   = Ignore;
#MaxInput = 0;

#Terminator = CR CR;
locktimeout = 5000;
replytimeout = 150;
readtimeout  = 80;
writetimeout = 120;
extrainput   = Ignore;
MaxInput = 0;

#terminator = CR LF;
#locktimeout = 10000;
#replytimeout = 5000;
#readtimeout  = 50;
#writetimeout = 100;
#extrainput   = Ignore;
#MaxInput = 0;



@init{out " ";}

# Get sensor information
sensors {
    out "Sensors";
	in  "Sensors %(\$1:\$2:\$3:SENSOR_INFO.VAL)s";
	in  "  State  SerialNumber   Name";
	in  "  %{InUse|Ready|Config|ERROR} %(\$1:\$2:\$3:SENSOR_SRLNO_I.VAL)s %(\$1:\$2:\$3:SENSOR_NAME_I.VAL)s";
}

# Select sensor information by Serial #
sensorsSelect {
	out "Select \$4";
#MANUAL	out "Select %s";
	in  "Select  OK";
	in  "  SerialNumber %(\$1:\$2:\$3:SELECTED_SRLNO_I.VAL)s";
  	in  "  State        %{InUse|Ready|Config|ERROR}";  
  	@mismatch {in "%(\$1:\$2:\$3:SENSOR_MESSAGE.VAL)s";}; 
}

read_current {out "CURRENT?"; in "CURRENT %f A"; @mismatch {in "%(\1.VAL)39c";}}

# Sensor State information  
sensorsState {
	out "SensorState";
	in  "SensorState OK"; 
  	in  "  State %{InUse|Ready|Config|ERROR}";
        in  "  UserApplication %(\$1:\$2:\$3:SENSOR_APP.VAL)s"; 
        in  "  UserVersion %(\$1:\$2:\$3:SENSOR_APP_VER.VAL)s";
        in  "  UserAddress %(\$1:\$2:\$3:SENSOR_IP.VAL)s"; 
#        @mismatch {in "%(\$1:\$2:\$3:Sensor\$3.VAL)s";}
}

control {
#	out 'Control "Procss Eye" "5.72"';
	out 'Control "EPICS"  "\$5"';
	in  "Control OK";
	in  "  SerialNumber  %(\$1:\$2:\$3:SELECTED_SRLNO_I.VAL)s";
}

release {
	out "Release";
	in  "Release OK";#%*s"; 
}

# Get the sensor available electonic gain factors 
egains {
	out "EGains";
	in  "EGains  OK";
	in  " %d";
}

# Retrieve current configuration and state of the RF Trip
rfInfo {
	out "RFInfo";
	in  "RFInfo  OK";
	in  "  RFTripEnabled %{No|Yes}";
	in  "  RFTripped %(\$1:\$2:\$3:RFTRIP_STS.VAL){No|Yes}";
}

# Get information about a specific set of source setting
sourceInfo {
	out "SourceInfo \$1";
	in  "SourceInfo  OK";
	in  "  SourceIndex         \$1";
	in  "  Name %s";
	in  "  ElectronEnergy %(\$2:\$3:\$4:SRC\$1_ENRG_I.VAL)f";
	in  "  IonEnergy %(\$2:\$3:\$4:SRC\$1_IONNRG_I.VAL)f";
	in  "  ExtractVolts %(\$2:\$3:\$4:SRC\$1_EXTRACTE_I.VAL)f";
	in  "  ElectronEmission %(\$2:\$3:\$4:SRC\$1_EEMI_I.VAL)f";
	in  "  LowMassAlignment %(\$2:\$3:\$4:SRC\$1_LOMASSALIGN_I.VAL)d";
	in  "  HighMassAlignment %(\$2:\$3:\$4:SRC\$1_HIMASSALIGN_I.VAL)d";
	in  "  LowMassResolution %(\$2:\$3:\$4:SRC\$1_LOMASSRES_I.VAL)d";
	in  "  HighMassResolution %(\$2:\$3:\$4:SRC\$1_HIMASSRES_I.VAL)d";
	in  "  MaxRecommendedPressure %(\$2:\$3:\$4:SRC\$1_MAXRMDP_I.VAL)e";
	in  "  NumFaradayEGains %(\$2:\$3:\$4:SRC\$1_NumFdyEGAIN_I.VAL)d";
	in  "  NumMultEGains %(\$2:\$3:\$4:SRC\$1_NumMultEGAIN_I.VAL)d";
	in  "  PoleBias %(\$2:\$3:\$4:SRC\$1_POLEBIAS_I.VAL)f";
}

# Get information about the detector settings for a particular source table
detectorInfo {
	out "DetectorInfo \$1";
	in  "DetectorInfo  OK";
	in  "  SourceIndex \$1";
	in  "  Name  DefaultFactor DefaultVoltage Factor1 Voltage1 Factor2 Voltage2 CalDate1 CalDate2";
	in  "  Faraday %e %(\$2:\$3:\$4:SRC\$1_FDYDVOLT_I.VAL)d %(\$2:\$3:\$4:SRC\$1_FDYFACT1_I.VAL)e %(\$2:\$3:\$4:SRC\$1_FDYVOLT1_I.VAL)f %(\$2:\$3:\$4:SRC\$1_FDYFACT2_I.VAL)e %(\$2:\$3:\$4:SRC\$1_FDYVOLT2_I.VAL)f %(\$2:\$3:\$4:SRC\$1_FDYCALDATE1_I.VAL)s %(\$2:\$3:\$4:SRC\$1_FDYCALDATE2_I.VAL)s";
	in  "  Mult1 %(\$2:\$3:\$4:SRC\$1_MULT1DFACT_I.VAL)e %(\$2:\$3:\$4:SRC\$1_MULT1DVOLT_I.VAL)d %(\$2:\$3:\$4:SRC\$1_MULT1FACT1_I.VAL)e %(\$2:\$3:\$4:SRC\$1_MULT1VOLT1_I.VAL)f %(\$2:\$3:\$4:SRC\$1_MULT1FACT2_I.VAL)e %(\$2:\$3:\$4:SRC\$1_MULT1VOLT2_I.VAL)f %(\$2:\$3:\$4:SRC\$1_MULT1CALDATE1_I.VAL)s %(\$2:\$3:\$4:SRC\$1_MULT1CALDATE2_I.VAL)s";
	in  "  Mult2 %(\$2:\$3:\$4:SRC\$1_MULT2DFACT_I.VAL)e %(\$2:\$3:\$4:SRC\$1_MULT2DVOLT_I.VAL)d %(\$2:\$3:\$4:SRC\$1_MULT2FACT1_I.VAL)e %(\$2:\$3:\$4:SRC\$1_MULT2VOLT1_I.VAL)f %(\$2:\$3:\$4:SRC\$1_MULT2FACT2_I.VAL)e %(\$2:\$3:\$4:SRC\$1_MULT2VOLT2_I.VAL)f %(\$2:\$3:\$4:SRC\$1_MULT2CALDATE1_I.VAL)s %(\$2:\$3:\$4:SRC\$1_MULT2CALDATE2_I.VAL)s";
	in  "  Mult3 %(\$2:\$3:\$4:SRC\$1_MULT3DFACT_I.VAL)e %(\$2:\$3:\$4:SRC\$1_MULT3DVOLT_I.VAL)d %(\$2:\$3:\$4:SRC\$1_MULT3FACT1_I.VAL)e %(\$2:\$3:\$4:SRC\$1_MULT3VOLT1_I.VAL)f %(\$2:\$3:\$4:SRC\$1_MULT3FACT2_I.VAL)e %(\$2:\$3:\$4:SRC\$1_MULT3VOLT2_I.VAL)f %(\$2:\$3:\$4:SRC\$1_MULT3CALDATE1_I.VAL)s %(\$2:\$3:\$4:SRC\$1_MULT3CALDATE2_I.VAL)s";	
}	

# Retrieve the current configuration and state of the filaments
filamentInfo {
	out "FilamentInfo";
	in  "FilamentInfo  OK";
	in  "  SummaryState  %{OFF|WARM-UP|ON|COOL-DOWN|BAD-EMISSION}";
	in  "  ActiveFilament %(\$1:\$2:\$3:ATVFIL_I.VAL)d";
	in  "  ExternalTripEnable %(\$1:\$2:\$3:EXTTRIPENA_STS.VAL){No|Yes}";
	in  "  ExternalTripMode  %(\$1:\$2:\$3:EXTTRIPMODE_STS.VAL){Trip|Control}";
	in  "  EmissionTripEnable %(\$1:\$2:\$3:EMITRIPENA_STS.VAL){No|Yes}";
	in  "  MaxOnTime %(\$1:\$2:\$3:MAXONK_I.VAL)d";
	in  "  OnTimeRemaining %(\$1:\$2:\$3:ONREMAINK_I.VAL)d";
	in  "  Trip  %(\$1:\$2:\$3:FILTRIP_STS.VAL){None|Emission|External|RVC|GaugeFail|TotalPressure}";
	in  "  Drive %(\$1:\$2:\$3:FILPS_STS.VAL){Off|On}";
	in  "  EmissionTripState %(\$1:\$2:\$3:EMITRIP_STS.VAL){Fail|OK}";
	in  "  ExternalTripState %(\$1:\$2:\$3:EXTTRIP_STS.VAL){Fail|OK}";
	in  "  RVCTripState  %(\$1:\$2:\$3:RVCTRIP_STS.VAL){Fail|OK}";
	in  "  GaugeTripState  %(\$1:\$2:\$3:GAUGETRIP_STS.VAL){Fail|OK}";

}

# Return information about all the analog inputs
aiInfo {
	out "AnalogInputInfo";
	in  "AnalogInputInfo OK";
	in  "  Enabled  MinVolts  MaxVolts  Resolution  Interval  AverageCount  Value";
	in  "  %{No|Yes}  %(\$1:\$2:\$3:MINVOLT_AI1_I.VAL)f %(\$1:\$2:\$3:MAXVOLT_AI1_I.VAL)f %(\$1:\$2:\$3:RES_AI1_I.VAL)f  %(\$1:\$2:\$3:INTV_AI1_I.VAL)d %(\$1:\$2:\$3:AVGCNT_AI1_I.VAL)d %(\$1:\$2:\$3:AI1_I.VAL)d";
	in  "  %(\$1:\$2:\$3:ENA_AI2_STS.VAL){No|Yes}  %(\$1:\$2:\$3:MINVOLT_AI2_I.VAL)f %(\$1:\$2:\$3:MAXVOLT_AI2_I.VAL)f %(\$1:\$2:\$3:RES_AI2_I.VAL)d  %(\$1:\$2:\$3:INTV_AI2_I.VAL)d %(\$1:\$2:\$3:AVGCNT_AI2_I.VAL)d %(\$1:\$2:\$3:AI2_I.VAL)d";
	in  "  %(\$1:\$2:\$3:ENA_AI3_STS.VAL){No|Yes}  %(\$1:\$2:\$3:MINVOLT_AI3_I.VAL)f %(\$1:\$2:\$3:MAXVOLT_AI3_I.VAL)f %(\$1:\$2:\$3:RES_AI3_I.VAL)d  %(\$1:\$2:\$3:INTV_AI3_I.VAL)d %(\$1:\$2:\$3:AVGCNT_AI3_I.VAL)d %(\$1:\$2:\$3:AI3_I.VAL)d";
	in  "  %(\$1:\$2:\$3:ENA_AI4_STS.VAL){No|Yes}  %(\$1:\$2:\$3:MINVOLT_AI4_I.VAL)f %(\$1:\$2:\$3:MAXVOLT_AI4_I.VAL)f %(\$1:\$2:\$3:RES_AI4_I.VAL)d  %(\$1:\$2:\$3:INTV_AI4_I.VAL)d %(\$1:\$2:\$3:AVGCNT_AI4_I.VAL)d %(\$1:\$2:\$3:AI4_I.VAL)d";
}

# Return information about all the analog outputs
aoInfo {
	out "AnalogOutputInfo";
	in  "AnalogOutputInfo  OK";
	in  "  MinVolts  MaxVolts  Resolution  Value";
	in  "  %f %(\$1:\$2:\$3:MAXVOLT_AO1_I.VAL)f %(\$1:\$2:\$3:RES_AO1_I.VAL)d %(\$1:\$2:\$3:AO1_I.VAL)d";
	in  "  %(\$1:\$2:\$3:MINVOLT_AO2_I.VAL)f %(\$1:\$2:\$3:MAXVOLT_AO2_I.VAL)f %(\$1:\$2:\$3:RES_AO2_I.VAL)d  %(\$1:\$2:\$3:AO2_I.VAL)d";
}

# Return information about the fitted digital input ports
digitalInfo {
	out "DigitalInfo";
	in  "DigitalInfo OK";
	in  "  DeglitchEnabled  %{No|Yes}";
	in  "  DeglitchTime %(\$1:\$2:\$3:DEGLITCHT_I.VAL)d";
	in  "  MaxPB67OnTime %(\$1:\$2:\$3:MAXPB67ONT_I.VAL)d";
	in  "  Name  ConnectedMask  OutputMask  Value";
	in  "  A %(\$1:\$2:\$3:CONNMASK_A_I.VAL)d %(\$1:\$2:\$3:OUTMASK_A_I.VAL)d %(\$1:\$2:\$3:DIG_A_I.VAL)d";
	in  "  B %(\$1:\$2:\$3:CONNMASK_B_I.VAL)d %(\$1:\$2:\$3:OUTMASK_B_I.VAL)d %(\$1:\$2:\$3:DIG_B_I.VAL)d";
}

# Return information about the mass program DAC and the resolution DAC
sourceTuneInfo {
	out "SourceTuningInfo";
	in  "SourceTuningInfo  OK";
	in  "  MassDACBits %d";
	in  "  MassDACUpperLimitM %(\$1:\$2:\$3:MASSDACHIM_I.VAL)f";
	in  "  MassDACUpperLimitC %(\$1:\$2:\$3:MASSDACHIC_I.VAL)f";
	in  "  MassDACLowerLimitM %(\$1:\$2:\$3:MASSDACLOM_I.VAL)f";
	in  "  MassDACLowerLimitC %(\$1:\$2:\$3:MASSDACLOC_I.VAL)f";
	in  "  ResolutionDACBits %(\$1:\$2:\$3:RESDACBITS_I.VAL)d";
}

# Return information about the available diagnostic input channels
diaginInfo {
	out "DiagnosticInputInfo";
	in  "DiagnosticInputInfo OK";
	in  "  Enabled Name                       Units       Interval AverageCount Value";
	in  "  %{No|Yes}     3.3V                       Volts %(\$1:\$2:\$3:INTER_3V_I.VAL)d %(\$1:\$2:\$3:AVGCNT_3V_I.VAL)d %(\$1:\$2:\$3:Diag3V_I.VAL)f";
	in  '  %(\$1:\$2:\$3:DIAG_RFIDENT_STS.VAL){No|Yes}      "RF Ident"                 Volts %(\$1:\$2:\$3:INTER_RFIDENT_I.VAL)d %(\$1:\$2:\$3:AVGCNT_RFIDENT_I.VAL)d %(\$1:\$2:\$3:DIAG_RFIDENT_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_EXTMON_STS.VAL){No|Yes}      "Extraction Monitor"       Volts %(\$1:\$2:\$3:INTER_EXTMON_I.VAL)d %(\$1:\$2:\$3:AVGCNT_EXTMON_I.VAL)d %(\$1:\$2:\$3:DIAG_EXTMON_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_NEG_15V_STS.VAL){No|Yes}     -15V                       Volts %(\$1:\$2:\$3:INTER_NEG_15V_I.VAL)d %(\$1:\$2:\$3:AVGCNT_NEG_15V_I.VAL)d %(\$1:\$2:\$3:DIAG_NEG_15V_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_RFK_STS.VAL){No|Yes}     "RF Temperature"           "Degrees C" %(\$1:\$2:\$3:INTER_RFK_I.VAL)d %(\$1:\$2:\$3:AVGCNT_RFK_I.VAL)d %(\$1:\$2:\$3:DIAG_RFK_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_RFTUNEE_STS.VAL){No|Yes}     "RF Tune Voltage"          Volts %(\$1:\$2:\$3:INTER_RFTUNEE_I.VAL)d %(\$1:\$2:\$3:AVGCNT_RFTUNEE_I.VAL)d %(\$1:\$2:\$3:DIAG_RFTUNEE_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_24V_STS.VAL){No|Yes}     +24V                       Volts %(\$1:\$2:\$3:INTER_24V_I.VAL)d %(\$1:\$2:\$3:AVGCNT_24V_I.VAL)d %(\$1:\$2:\$3:Diag24V_I.VAL)f'; 
	in  '  %(\$1:\$2:\$3:DIAG_NEG_130V_STS.VAL){No|Yes}     -130V                      Volts %(\$1:\$2:\$3:INTER_NEG_130V_I.VAL)d %(\$1:\$2:\$3:AVGCNTNEG_130V_I.VAL)d %(\$1:\$2:\$3:DIAG_NEG_130V_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_IONNRG_STS.VAL){No|Yes}      "Ion Energy"               Volts %(\$1:\$2:\$3:INTER_IONNRG_I.VAL)d %(\$1:\$2:\$3:AVGCNT_IONNRG_I.VAL)d %(\$1:\$2:\$3:DIAG_IONNRG_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_SEMMON_STS.VAL){No|Yes}      "SEM Monitor"              Volts %(\$1:\$2:\$3:INTER_SEMMON_I.VAL)d %(\$1:\$2:\$3:AVGCNT_SEMMON_I.VAL)d %(\$1:\$2:\$3:DIAG_SEMMON_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_SRCIMON_STS.VAL){No|Yes}      "Source Current Monitor"   mA %(\$1:\$2:\$3:INTER_SRCIMON_I.VAL)d %(\$1:\$2:\$3:AVGCNT_SRCIMON_I.VAL)d %(\$1:\$2:\$3:DIAG_SRCIMON_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_MASSDAC_STS.VAL){No|Yes}      "Mass DAC"                 Volts %(\$1:\$2:\$3:INTER_MASSDAC_I.VAL)d %(\$1:\$2:\$3:AVGCNT_MASSDAC_I.VAL)d %(\$1:\$2:\$3:DIAG_MASSDAC_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_FILI_STS.VAL){No|Yes}      "Filament Current"         Amps %(\$1:\$2:\$3:INTER_FILI_I.VAL)d %(\$1:\$2:\$3:AVGCNT_FILI_I.VAL)d %(\$1:\$2:\$3:DIAG_FILI_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_ENRG_STS.VAL){No|Yes}      "Electron Energy"          Volts %(\$1:\$2:\$3:INTER_ENRG_I.VAL)d %(\$1:\$2:\$3:AVGCNT_ENRG_I.VAL)d %(\$1:\$2:\$3:DIAG_ENRG_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_RES_STS.VAL){No|Yes}      Resolution                 Volts %(\$1:\$2:\$3:INTER_RES_I.VAL)d %(\$1:\$2:\$3:AVGCNT_RES_I.VAL)d %(\$1:\$2:\$3:DIAG_RES_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_POLEBIAS_STS.VAL){No|Yes}     "Pole Bias"                Volts %(\$1:\$2:\$3:INTER_POLEBIAS_I.VAL)d %(\$1:\$2:\$3:AVGCNT_POLEBIAS_I.VAL)d %(\$1:\$2:\$3:DIAG_POLEBIAS_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_EMETERK_STS.VAL){No|Yes}     "Electrometer Temperature" "Degrees C" %(\$1:\$2:\$3:INTER_EMETERK_I.VAL)d %(\$1:\$2:\$3:AVGCNT_EMETERK_I.VAL)d %(\$1:\$2:\$3:DIAG_EMETERK_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_15V_STS.VAL){No|Yes}     +15V                       Volts %(\$1:\$2:\$3:INTER_15V_I.VAL)d %(\$1:\$2:\$3:AVGCNT_15V_I.VAL)d %(\$1:\$2:\$3:Diag15V_I.VAL)f';
	in  '  %(\$1:\$2:\$3:DIAG_INTK_STS.VAL){No|Yes}     "Internal Temperature"     "Degrees C" %(\$1:\$2:\$3:INTER_INTK_I.VAL)d %(\$1:\$2:\$3:AVGCNT_INTK_I.VAL)d %(\$1:\$2:\$3:DIAG_INTK_I.VAL)f';
}
	


filamentControl {
	out "FilamentControl %{Off|On}";
	in  "FilamentControl OK";
	in  "  State %{Off|On}";
}

filamentSelect
{
	out "FilamentSelect %{1|2}"; 
	in  "FilamentSelect  OK";
	in  "  Number %*d";
}

filamentOnT {
	out "FilamentOnTime %d";
	in  "FilamentOnTime  OK";
	in  "  Time %*d";
}

addAnalog {
	out "AddAnalog Analog1 %(\$1:\$2:\$3:ADDANL_STRTMASS_SP.VAL)d %(\$1:\$2:\$3:ADDANL_ENDMASS_SP.VAL)d %(\$1:\$2:\$3:ADDANL_POINTS_PK_SEL.VAL){1|2|4|8|16|32}  %(\$1:\$2:\$3:AddAnalog:ACCURACY_SEL.VAL)d %(\$1:\$2:\$3:ADDANL_EGAIN_INDX_SEL.VAL)d %(\$1:\$2:\$3:ADDANL_SRCINDX_SEL.VAL)d %(\$1:\$2:\$3:ADDANL_DETCINDX_SEL.VAL)d";
	in  "AddAnalog OK";
	in  "  Name %*s";
	in  "  StartMass %*d";
 	in  "  EndMass %*d";
	in  "  PointsPerPeak %*d";
	in  "  Accuracy %*d";
	in  "  EGainIndex %*d";
	in  "  SourceIndex %*d";
	in  "  DetectorIndex %*d";
}

addBarchart {
	out "AddBarchart Bar1 %d %(\$1:\$2:\$3:ADDBAR_ENDMASS_SP.VAL)d %(\$1:\$2:\$3:ADDBAR_FILTERMODE_SEL.VAL){PeakCenter|PeakMax|PeakAverage}  %(\$1:\$2:\$3:AddBar:ACCURACY_SEL.VAL)d %(\$1:\$2:\$3:ADDBAR_EGAININDX_SEL.VAL)d %(\$1:\$2:\$3:ADDBAR_SRCINDX_SEL.VAL)d %(\$1:\$2:\$3:ADDBAR_DETCINDX_SEL.VAL)d";
	in  "AddBarchart OK";
	in  "  Name                   Bar1";
	in  "  StartMass %*d";
 	in  "  EndMass %*d";
	in  "  FilterMode %*s";
	in  "  Accuracy %*d";
	in  "  EGainIndex %*d";
	in  "  SourceIndex %*d";
	in  "  DetectorIndex %*d";
}

addPeakjump {
	out "AddPeakJump %s %(\$1:\$2:\$3:ADDPKJUMP_FLTRMODE.VAL){PeakCenter|PeakMax|PeakAverage}  %(\$1:\$2:\$3:AddPeakJump:ACCURACY_SEL.VAL)d %(\$1:\$2:\$3:ADDPKJUMP_EGAININDX.VAL)d %(\$1:\$2:\$3:ADDPKJUMP_SRCINDX_SEL.VAL)d %(\$1:\$2:\$3:ADDPKJUMP_DETCINDX_SEL.VAL)d";
	in  "AddPeakJump OK";
	in  "  Name %*s";
	in  "  FilterMode %*s";
	in  "  Accuracy %*d";
	in  "  EGainIndex %*d";
	in  "  SourceIndex %*d";
	in  "  DetectorIndex %*d";
}

addSinglepeak {
	out "AddSinglePeak %s %(\$1:\$2:\$3:ADDSINGLEPK_MASS_SP.VAL)d  %(\$1:\$2:\$3:AddSinglePeak:ACCURACY_SEL.VAL)d %(\$1:\$2:\$3:ADDSING_PK_EGAININDX.VAL)d %(\$1:\$2:\$3:ADDSINGLEPK_SRCINDX.VAL)d %(\$1:\$2:\$3:ADDSINGLEPK_DETCINDX.VAL)d";
	in  "AddSinglePeak OK";
	in  "  Name %*s";
	in  "  Mass %*d";
	in  "  Accuracy %*d";
	in  "  EGainIndex %*d";
	in  "  SourceIndex %*d";
	in  "  DetectorIndex %*d";
}

measAccuracy {
	out "MeasurementAccuracy %{0|1|2|3|4|5|6|7|8}";
	in  "MeasurementAccuracy OK";
	in  "  Accuracy  %*d";
}

measAddMass {
	out "MeasurementAddMass %d";
	in  "MeasurementAddMass  OK";
	in  "  Mass %*d";
}

measChangeMass {
	out "MeasurementChangeMass %d %(\$1:\$2:\$3:MEASCHMASS_INDX_SP.VAL)d";
	in  "MeasurementChangeMass OK";
	in  "  MassIndex %*d";
	in  "  NewMass %*d";
}

measDetectorIndex {
	out "MeasurementDetectorIndex %d";
	in  "MeasurementDetectorIndex OK";
	in  "  DetectorIndex %*d";
}

measEGainIndex {
	out "MeasurementEGainIndex %d";
	in  "MeasurementEGainIndex OK";
	in  "  EGainIndex %*d";
}

measFilterMode {
	out "MeasurementFilterMode %{PeakCenter|PeakMax|PeakAverage}";
	in  "MeasurementFilterMode OK";
	in  "  FilterMode %*s";
}

measMass {
	out "MeasurementMass %f";
	in  "MeasurementMass OK";
	in  "  Mass %*f";
}

measPointsPerPeak {
	out "MeasurementPointsPerPeak %{1|2|4|8|16|32}";
	in  "MeasurementPointsPerPeak OK";
	in  "  PointsPerPeak %*d";
}

measRemoveMass {
	out "MeasurementRemoveMass %d";
	in  "MeasurementRemoveMass OK";
	in  "  MassIndex %*d";
}

measSourceIndex {
	out "MeasurementSourceIndex %d";
	in  "MeasurementSourceIndex OK";
	in  "  SourceIndex %*d";
}

measZeroBeamOff {
	out "MeasurementZeroBeamOff %{False|True}";
	in  "MeasurementZeroBeamOff OK";
	in  "  BeamOff %*s";
}

measZeroBufferDepth {
	out "MeasurementZeroBufferDepth %{1|2|4|8|16|32}";
	in  "MeasurementZeroBufferDepth OK";
	in  "  ZeroBufferDepth %*d";
}

measZeroBufferMode {
	out "MeasurementZeroBufferMode %{SingleScanAverage|MultiScanAverage|MultiScanAverageQuickStart|SingleShot}";
	in  "MeasurementZeroBufferMode OK";
	in  "  ZeroBufferMode %*s";
}

measZeroRetrigger {
	out "MeasurementZeroReTrigger";
	in  "MeasurementZeroReTrigger OK";
}

measZeroMass {
	out "MeasurementZeroMass %f";
	in  "MeasurementZeroMass OK";
	in  "  ZeroMass %*f";
}

multiplierProtect {
	out "MultiplierProtect %{False|True}";
	in  "MultiplierProtect OK";
}

runDiagnostics {
	out "RunDiagnostics";
	in  "RunDiagnostics  OK";
	in  "  Name                       Min      Max      Value    Result";
	in  "  3.3V %(\$1:\$2:\$3:3V_I.LOW)f %(\$1:\$2:\$3:3V_I.HIGH)f %f %(\$1:\$2:\$3:RESULT_3V_STS.VAL){N/A|Fail|Pass}";
	in  "  +15V %(\$1:\$2:\$3:15V_I.LOW)f %(\$1:\$2:\$3:15V_I.HIGH)f %(\$1:\$2:\$3:15V_I.VAL)f %(\$1:\$2:\$3:RESULT_15V_STS.VAL){N/A|Fail|Pass}";
	in  "  -15V %(\$1:\$2:\$3:NEG_15V_I.LOW)f %(\$1:\$2:\$3:NEG_15V_I.HIGH)f %(\$1:\$2:\$3:NEG_15V_I.VAL)f %(\$1:\$2:\$3:RESULT_NEG_15V_STS.VAL){N/A|Fail|Pass}";
	in  "  +24V %(\$1:\$2:\$3:24V_I.LOW)f %(\$1:\$2:\$3:24V_I.HIGH)f %(\$1:\$2:\$3:24V_I.VAL)f %(\$1:\$2:\$3:RESULT_24V_STS.VAL){N/A|Fail|Pass}";
	in  "  -130V %(\$1:\$2:\$3:NEG_130V_I.LOW)f %(\$1:\$2:\$3:NEG_130V_I.HIGH)f %(\$1:\$2:\$3:NEG_130V_I.VAL)f %(\$1:\$2:\$3:RESULT_NEG_130V_STS.VAL){N/A|Fail|Pass}";
}

setTotalP {
	out "SetTotalPressure %6.1e";
	in  "SetTotalPressure  OK";
}

calibrationOpt {
	out "CalibrationOptions %{Off|Default|Current} %(\$1:\$2:\$3:CALOPT_DETC_SEL.VAL){Off|Default|Current}";
	in  "CalibrationOptionss   OK";
}

detectorFactor {
	out "DetectorFactor %(\$1:\$2:\$3:DETCFACTOR_SRCINDX_SEL.VAL)d %(\$1:\$2:\$3:DETCFACTOR_DETCINDX.VAL)d %(\$1:\$2:\$3:DETCFACTOR_FIL_SEL.VAL){1|2} %6.1e";
	in  "DetectorFactor  OK";
}

detectorCalDate {
	out "DetectorCalDate %(\$1:\$2:\$3:DETCFACTOR_SRCINDX_SEL.VAL)d %(\$1:\$2:\$3:DETCFACTOR_DETCINDX.VAL)d %(\$1:\$2:\$3:DETCFACTOR_FIL_SEL.VAL){1|2} %s";
	in  "DetectorCalDate OK";
}

detectorVoltage {
	out "DetectorVoltage %(\$1:\$2:\$3:DETCFACTOR_SRCINDX_SEL.VAL)d %(\$1:\$2:\$3:DETCFACTOR_DETCINDX.VAL)d %(\$1:\$2:\$3:DETCFACTOR_FIL_SEL.VAL){1|2} %f";
	in  "DetectorVoltage OK";
}

# Add a measurement to the scans list of measurements
scanAdd {
	out "ScanAdd %{Analog1|Bar1}";
	in  "ScanAdd OK";
	in  "  Measurement %*s";
}
#add bar scan
scanAddBar {
	out "ScanAdd Bar1";
	in  "ScanAdd OK";
	in  "  Measurement %*s";
}
#add analog scan
scanAddAnalog {
	out "ScanAdd Analog1";
	in  "ScanAdd OK";
	in  "  Measurement %*s";
}
# Start a scan running and will re-trigger the scan automatically the number of times specified by number
scanStart {
	out "ScanStart %(\$1:\$2:\$3:SCANNUM_SP.VAL)d";
	in  "ScanStart OK";
}

getMeasT {
	in  "StartingScan %*d %d %*d";
}

getRFTripStatus {
	in  "RFTripState %{Tripped|OK}";
}

activeMeas {
	in "StartingMeasurement %?#{\$1=1}";
}

getMeasZero {
	in  "ZeroReading %(\$1:\$2:\$3:ZERO_MASS_I.VAL)f %e";
}

getMeasVal {
	in  "MassReading %(\$1:\$2:\$3:CURMASS_I.VAL)f %e";
}

# Stop a scan running and remove all measurements from the scan list
scanStop {
	out "SCANSTOP";
	in  "SCANSTOP  OK";
}

# Re-trigger the scan specified by number
scanResume {
	out "ScanResume %(\$1:\$2:\$3:SCANNUM_SP.VAL)d";
	in  "ScanResume  OK";
}


# Re-start the current scan from the beginning
scanRestart {
	out "ScanRestart";
	in  "ScanRestart OK";
}

# Select a measurement to act on
measSelect {
	out "MeasurementSelect %s";
	in  "MeasurementSelect OK";
	in  "  Measurement %*s";
}

# Set selected analog or barchart measurements starting mass
measStartMass {
	out "MeasurementStartMass %d";
	in  "MeasurementStartMass OK";
	in  "  StartMass %*d";
	in  "  EndMass %*d";
}

# Set selected analog or barchart measurements ending mass
measEndMass {
	out "MeasurementEndMass %d";
	in  "MeasurementEndMass  OK";
	in  "  StartMass %*d";
	in  "  EndMass %*d";
}

# Set number of scans of consecutive full scale readings before skipping mass. 0 to disable
measMultSkipAutoPro {
	out "MeasurementMultSkipAutoProtect %d";
	in  "MeasurementMultSkipAutoProtect OK";
	in  "  ProtectScans %*d";
}

# Add a mass to the list of masses to be skipped by the selected measurement when multiplier is being used
measMultSkipMassAdd {
	out "MeasurementMultSkipMassAdd %d";
	in  "MeasurementMultSkipMassAdd OK";
	in  "  Mass %*d";
}

# Remove a mass from the multiplier skip list so that it will be scanned again with the multiplier detector
measMultSkipMassRemove {
	out "MeasurementMultSkipMassRemove %d";
	in  "MeasurementMultSkipMassRemove OK";
	in  "  Mass %*d";
}

# Remove all masses from the multiplier skip list so that they will again be scanned with the multiplier detector
measMultSkipMassRemoveAll {
	out "MeasurementMultSkipMassRemoveAll";
	in  "MeasurementMultSkipMassRemoveAll OK";
}

# Remove all measurements from the sensor
measRemoveAll {
	out "MeasurementRemoveAll";
	in  "MeasurementRemoveAll OK";
}

# Remove the specified measurement from the sensor
measRemove {
	out "MeasurementRemove %s";
	in  "MeasurementRemove OK";
	in  "  Measurement %*s";
}

# Set a source settings parameters ion energy setting
sourceIonEnergy {
	out "SourceIonEnergy %(\$1:\$2:\$3:SRCINDX_SEL.VAL)d %f";
	in  "SourceIonEnergy OK";
}

# Set a source settings parameters emission setting
sourceEmission {
	out "SourceEmission %(\$1:\$2:\$3:SRCINDX_SEL.VAL)d %f";
	in  "SourceEmission  OK";
}

# Set a source settings parameters extract setting
sourceExtract {
	out "SourceExtract %(\$1:\$2:\$3:SRCINDX_SEL.VAL)d %f";
	in  "SourceExtract OK";
}

# Set a source settings parameters Electron energy setting
sourceENrg {
	out "SourceElectronEnergy %(\$1:\$2:\$3:SRCINDX_SEL.VAL)d %f";
	in  "SourceElectronEnergy OK";
}

# Set a source settings parameters low mass resolution setting
#sourceLowMassRes {
#	out "SourceLowMassResolution %(\$1:\$2:\$3:SRCINDX_SEL.VAL)d %d";
#	in  "SourceLowMassResolution OK";
#}

# Set a source settings parameters low mass alignment setting
#sourceLowMassAlign {
#	out "SourceLowMassAlignment %(\$1:\$2:\$3:SRCINDX_SEL.VAL)d %d";
#	in  "SourceLowMassAlignment OK";
#}

# Set a source settings parameters high mass resolution setting
#sourceHighMassRes {
#	out "SourceHighMassResolution %(\$1:\$2:\$3:SRCINDX_SEL.VAL)d %d";
#	in  "SourceHighMassResolution OK";
#}

# Set a source settings parameters high mass alignment setting
#sourceHighMassAlign {
#	out "SourceHighMassAlignment %(\$1:\$2:\$3:SRCINDX_SEL.VAL)d %d";
#	in  "SourceHighMassAlignment OK";
#}

# Set the number of readings that should be taken and averaged before results are sent back
aiAvgCnt {
	out "AnalogInputAverageCount %(\$1:\$2:\$3:INDX_AI_SP.VAL)d %d";
	in  "AnalogInputAverageCount OK";
}

# Enable or disable analog input readings from being sent when in control of the sensor
aiEnable {
	out "AnalogInputEnable %(\$1:\$2:\$3:INDX_AI_SP.VAL)d %{False|True}";
	in  "AnalogInputEnable OK";
}

# Set the interval between analog input readings in the sensor
aiInterval {
	out "AnalogInputInterval %(\$1:\$2:\$3:INDX_AI_SP.VAL)d %d";
	in  "AnalogInputInterval OK";
}

# Set a given analog output channel to the specified voltage
aoVal {
	out "AnalogOutput %(\$1:\$2:\$3:INDX_AO_SP.VAL)d %f";
	in  "AnalogOutput  OK";
}

# Set audio frequency
audioFrequency {
	out "AudioFrequency %d";
	in  "AudioFrequency  OK";
}

# Set audio mode
audioMode {
	out "AudioMode %{Off|Automatic|Manual}";
	in  "AudioMode OK";
}

# Set the time that either pin 6 or 7 will remain set for after they're initially set
digitalMaxPB67OnT {
	out "DigitalMaxPB67OnTime %d";
	in  "DigitalMaxPB67OnTime OK";
}

# Set digital outputs according to the value specified
digitalOut {
	out "DigitalOutput \$1 %d";
	in  "DigitalOutput OK";
} 

# Save changes
saveChanges {
	out "SaveChanges";
	in  "SaveChanges OK";
}

# Run a degas operation
startDegas {
	out "StartDegas %(\$1:\$2:\$3:STRTDEGAS_STRTPWR_SP.VAL)d %(\$1:\$2:\$3:STRTDEGAS_ENDPWR_SP.VAL)d %(\$1:\$2:\$3:STRTDEGAS_RAMP_PRD_SP.VAL)d %(\$1:\$2:\$3:STRTDEGAS_MAXPWR_PRD_SP.VAL)d %(\$1:\$2:\$3:STRT_DEGAS_SETTLE_PRD_SP.VAL)d";
	in  "StartDegas  OK";
}

# End a degas operation
stopDegas {
	out "StopDegas";
	in  "StopDegas OK";
}

# Copy the specified source tables alignement data to all other source tables
sourceAlignCopyToAll {
	out "SourceAlignmentCopyToAll %(\$1:\$2:\$3:SRCALIGN_SRCINDX_SEL.VAL)d";
	in  "SourceAlignmentCopyToAll OK";
}

# Remove a specified mass from mass alignement table for the specified source table
sourceAlignRemove {
	out "SourceAlignmentRemove %(\$1:\$2:\$3:SRCALIGN_SRCINDX_SEL.VAL)d %d";
	in  "SourceAlignmentRemove OK";
}

# Copy the specified source tables resolution data to all other source tables
sourceResCopyToAll {
	out "SourceResolutionCopyToAll %(\$1:\$2:\$3:SourceRes:SRCINDX_SEL.VAL)d";
	in  "SourceResolutionCopyToAll OK";
}

# Remove a specified mass from mass resolution table for the specified source table
sourceResRemove {
	out "SourceResolutionRemove %(\$1:\$2:\$3:SourceRes:SRCINDX_SEL.VAL)d %d";
	in  "SourceResolutionRemove OK";
}

# Set pole bias settings to a specified source table entry
sourcePoleBias {
	out "SourcePoleBias %(\$1:\$2:\$3:SourcePoleBias:SRCINDX_SEL.VAL)d %f";
	in  "SourcePoleBias  OK";
}

# Set the number of readings that should be taken and averaged before results are sent back
diagInAvgCnt {
	out "DiagnosticInputAverageCount %(\$1:\$2:\$3:AIINDX_SP.VAL)d %d";
	in  "DiagnosticInputAverageCount OK";
}

# Enable or disable diagnostic input readings from being sent when in control of the sensor
diagInEnable {
	out "DiagnosticInputEnable %(\$1:\$2:\$3:AIINDX_SP.VAL)d %{False|True}";
	in  "DiagnosticInputEnable OK";
}

# Set the interval between diagnostic input readings in the sensor
diagInInterval {
	out "DiagnosticInputInterval %(\$1:\$2:\$3:AIINDX_SP.VAL)d %d";
	in  "DiagnosticInputInterval OK";
}