# agilent_3458a.proto
# 20210527 mdunning
#
# This device is really slow, so the timeouts below are actually needed.
# Fetching data is especially slow, and can take ~60 seconds or longer.

InTerminator = CR;
OutTerminator = CR LF;

ReadTimeout = 5000;
WriteTimeout = 60000;
ReplyTimeout = 5000;
LockTimeout = 5000;

getString {
    out "\$1";
    in "%#s";
}

setString {
    out "\$1";
    @mismatch {
        in "";
    }
}

getInt {
    out "\$1";
    in "%d";
    @mismatch {
        in "";
    }
}

setInt {
    out "\$1 %d";
    @mismatch {
        in "";
    }
}

getFloat {
    out "\$1";
    in "%f";
}

setFloat {
    out "\$1 %f";
    @mismatch {
        in "";
    }
}

getMeasMode {
    out "\$1";
    in "%d,%(\$2:MeasRange_RBV)f";
}

setMeasMode {
    out "\$1 %(\$2:MeasMode)d,%(\$2:MeasRange)f,%(\$2:MeasRes)f";
    @mismatch {
        in "";
    }
}

getMathMode {
    out "\$1";
    in "%d,%(\$2:MathMode2_RBV)d";
}

setMathMode {
    out "\$1 %(\$2:MathMode1)d,%(\$2:MathMode2)d";
    @mismatch {
        in "";
    }
}

getNumReadings {
    out "\$1";
    in "%(\$2:NumReadings_RBV)d,%d";
}

setNumReadings {
    out "\$1 %(\$2:NumReadings)d,%d";
    @mismatch {
        in "";
    }
}

getMem {
    out "RMEM 1,%d";
    separator = ",";
    in "%(\$1)f";
}

setTrig {
    out "\$1 %d";
    @mismatch {
        in "%(\$2:ValStr)s";
    }
}

setTrigGetData {
    out "\$1 %d";
    in "%(\$2:MeasValue)f";
    @mismatch {
        in "%(\$2:ValStr)s";
    }
}