# agilent_3458a.db
# 20210527 md
#
# Since this device is so slow, "automatic" readbacks (FLNKs to RBV records) are disabled.
# Periodic scanning of records is also disabled.

record(stringin, "$(P):ID") {
    field(DESC, "Device ID")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getString(ID?) $(PORT) $(ADDR)")
}

record(longin, "$(P):Error") {
    field(DESC, "Error status")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(ERR?) $(PORT) $(ADDR)")
    field(HIHI, "1")
    field(HHSV, "MAJOR")
}

record(mbbiDirect, "$(P):Status") {
    field(DESC, "Status byte")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(STB?) $(PORT) $(ADDR)")
}

record(bo, "$(P):Reset") {
    field(DESC, "Reset device")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setString(RESET) $(PORT) $(ADDR)")
    field(ZNAM, "Standby")
    field(ONAM, "Reset")
}

record(bo, "$(P):Preset") {
    field(DESC, "Set to preset state")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setString(PRESET) $(PORT) $(ADDR)")
    field(ZNAM, "Standby")
    field(ONAM, "Preset")
}

record(ai, "$(P):MeasValue") {
    field(DESC, "Measurement value")
    field(PREC, "6")
}

record(stringin, "$(P):ValStr") {
    field(DESC, "Value string")
}


record(bo, "$(P):LineFreq") {
    field(DESC, "Line freq")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setString(LFREQ LINE) $(PORT) $(ADDR)")
    field(ZNAM, "Standby")
    field(ONAM, "Set")
#    field(FLNK, "$(P):LineFreq_RBV")
}

record(ai, "$(P):LineFreq_RBV") {
    field(DESC, "Line freq")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getFloat(LFREQ?) $(PORT) $(ADDR)")
    field(PREC, "6")
    field(EGU,  "Hz")
}


record(ao, "$(P):IntTime") {
    field(DESC, "Int time: num pwr line cyc")
    field(DTYP, "stream")
    field(LOPR, "0")
    field(HOPR, "1000")
    field(DRVL, "0")
    field(DRVH, "1000")
    field(EGU,  "NPLC")
    field(PREC, "6")
    field(OUT,  "@$(PROTOFILE) setFloat(NPLC, $(P)) $(PORT) $(ADDR)")
#    field(FLNK, "$(P):IntTime_RBV")
}

record(ai, "$(P):IntTime_RBV") {
    field(DESC, "Int time: num pwr line cyc")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getFloat(NPLC?) $(PORT) $(ADDR)")
    field(EGU,  "NPLC")
    field(PREC, "6")
}


record(mbbo, "$(P):OutFormat") {
    field(DESC, "Outset format")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setInt(OFORMAT, $(P)) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "")
    field(ONVL, "1")
    field(ONST, "ASCII")
    field(TWVL, "2")
    field(TWST, "SINT")
    field(THVL, "3")
    field(THST, "DINT")
    field(FRVL, "4")
    field(FRST, "SREAL")
    field(FVVL, "5")
    field(FVST, "DREAL")
#    field(FLNK, "$(P):OutFormat_RBV")
}

record(mbbi, "$(P):OutFormat_RBV") {
    field(DESC, "Outset format")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(OFORMAT?) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "")
    field(ONVL, "1")
    field(ONST, "ASCII")
    field(TWVL, "2")
    field(TWST, "SINT")
    field(THVL, "3")
    field(FRVL, "4")
    field(FRST, "SREAL")
    field(FVVL, "5")
    field(FVST, "DREAL")
    field(THST, "DINT")
}


record(mbbo, "$(P):TrigMode") {
    field(DESC, "Trigger mode")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setTrig(TRIG, $(P)) $(PORT) $(ADDR)")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "Auto")
    field(TWVL, "2")
    field(TWST, "Ext")
    field(THVL, "3")
    field(THST, "SGL")
    field(FRVL, "4")
    field(FRST, "HOLD")
    field(FVVL, "5")
    field(FVST, "SYN")
    field(SVVL, "7")
    field(SVST, "Level")
    field(EIVL, "8")
    field(EIST, "Line")
#    field(FLNK, "$(P):TrigMode_RBV")
}

record(mbbi, "$(P):TrigMode_RBV") {
    field(DESC, "Outset format")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(TRIG?) $(PORT) $(ADDR)")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "Auto")
    field(TWVL, "2")
    field(TWST, "Ext")
    field(THVL, "3")
    field(THST, "SGL")
    field(FRVL, "4")
    field(FRST, "HOLD")
    field(FVVL, "5")
    field(FVST, "SYN")
    field(SVVL, "7")
    field(SVST, "Level")
    field(EIVL, "8")
    field(EIST, "Line")
}

record(mbbo, "$(P):TrigData") {
    field(DESC, "Trigger and get data")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setTrigGetData(TRIG, $(P)) $(PORT) $(ADDR)")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "Auto")
    field(TWVL, "2")
    field(TWST, "Ext")
    field(THVL, "3")
    field(THST, "SGL")
    field(FRVL, "4")
    field(FRST, "HOLD")
    field(FVVL, "5")
    field(FVST, "SYN")
    field(SVVL, "7")
    field(SVST, "Level")
    field(EIVL, "8")
    field(EIST, "Line")
}


record(ai, "$(P):IntegerScale_RBV") {
    field(DESC, "Integer scale")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getFloat(ISCALE?) $(PORT) $(ADDR)")
    field(PREC, "6")
    field(EGU,  "")
}

#--- Measurement mode --------
record(mbbo, "$(P):MeasMode") {
    field(DESC, "Measurement mode")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "DCV")
    field(TWVL, "2")
    field(TWST, "ACV")
    field(THVL, "3")
    field(THST, "ACDCV")
    field(FRVL, "4")
    field(FRST, "OHM")
    field(FVVL, "5")
    field(FVST, "OHMF")
    # Six state does not exist
    field(SVVL, "7")
    field(SVST, "ACI")
    field(EIVL, "8")
    field(EIST, "ACDCI")
    field(NIVL, "9")
    field(NIST, "FREQ")
    field(TEVL, "10")
    field(TEST, "PER")
    field(ELVL, "11")
    field(ELST, "DSAC")
    field(TVVL, "12")
    field(TVST, "DSDC")
    field(TTVL, "13")
    field(TTST, "SSAC")
    field(FTVL, "14")
    field(FTST, "SSDC")
    field(PINI, "1")
    info(autosaveFields, "VAL")
}

record(mbbi, "$(P):MeasMode_RBV") {
    field(DESC, "Measurement mode")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getMeasMode(FUNC?, $(P)) $(PORT) $(ADDR)")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "DCV")
    field(TWVL, "2")
    field(TWST, "ACV")
    field(THVL, "3")
    field(THST, "ACDCV")
    field(FRVL, "4")
    field(FRST, "OHM")
    field(FVVL, "5")
    field(FVST, "OHMF")
    # Six state does not exist
    field(SVVL, "7")
    field(SVST, "ACI")
    field(EIVL, "8")
    field(EIST, "ACDCI")
    field(NIVL, "9")
    field(NIST, "FREQ")
    field(TEVL, "10")
    field(TEST, "PER")
    field(ELVL, "11")
    field(ELST, "DSAC")
    field(TVVL, "12")
    field(TVST, "DSDC")
    field(TTVL, "13")
    field(TTST, "SSAC")
    field(FTVL, "14")
    field(FTST, "SSDC")
}

record(ao, "$(P):MeasRange") {
    field(DESC, "Measurement range")
    field(PREC, "6")
    field(EGU,  "")
    field(PINI, "1")
    info(autosaveFields, "VAL PREC")
}

record(ai, "$(P):MeasRange_RBV") {
    field(DESC, "Measurement range")
    field(PREC, "2")
    field(EGU,  "")
}

record(ao, "$(P):MeasRes") {
    field(DESC, "Measurement resolution")
    field(PREC, "6")
    field(EGU,  "%")
    field(PINI, "1")
    info(autosaveFields, "VAL PREC")
}

# This record writes the mode, range, and resolution to the device.
# The user should set the MeasMode, MeasRange, and MeasRes PVs first, then 
#   process this record.
record(bo, "$(P):MeasSet") {
    field(DESC, "Measurement set")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setMeasMode(FUNC, $(P)) $(PORT) $(ADDR)")
    field(ZNAM, "Standby")
    field(ONAM, "Set")
#    field(FLNK, "$(P):MeasMode_RBV")
}
#--- End Measurement mode --------

record(mbbo, "$(P):AutorangeMode") {
    field(DESC, "Autorange mode")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setInt(ARANGE, $(P)) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "ON")
    field(TWVL, "2")
    field(TWST, "ONCE")
#    field(FLNK, "$(P):AutorangeMode_RBV")
}

record(mbbi, "$(P):AutorangeMode_RBV") {
    field(DESC, "Autorange mode")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(ARANGE?) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "ON")
    field(TWVL, "2")
    field(TWST, "ONCE")
}


record(mbbo, "$(P):AutozeroMode") {
    field(DESC, "Autozero mode")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setInt(AZERO, $(P)) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "ON")
    field(TWVL, "2")
    field(TWST, "ONCE")
#    field(FLNK, "$(P):AutozeroMode_RBV")
}

record(mbbi, "$(P):AutozeroMode_RBV") {
    field(DESC, "Autozero mode")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(AZERO?) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "ON")
    field(TWVL, "2")
    field(TWST, "ONCE")
}


record(mbbo, "$(P):DisplayMode") {
    field(DESC, "Display mode")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setInt(DISP, $(P)) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "ON")
    field(TWVL, "2")
    field(TWST, "MSG")
    field(THVL, "3")
    field(THST, "CLR")
#    field(FLNK, "$(P):DisplayMode_RBV")
}

record(mbbi, "$(P):DisplayMode_RBV") {
    field(DESC, "Display mode")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(DISP?) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "ON")
    field(TWVL, "2")
    field(TWST, "MSG")
    field(THVL, "3")
    field(THST, "CLR")
}


#--- Math mode --------
record(mbbo, "$(P):MathMode1") {
    field(DESC, "Math mode 1")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setMathMode(MATH, $(P)) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "CONT")
    # Two state does not exist
    field(THVL, "3")
    field(THST, "CTHRM")
    field(FRVL, "4")
    field(FRST, "DB")
    field(FVVL, "5")
    field(FVST, "DBM")
    field(SXVL, "6")
    field(SXST, "FILTER")
    # Seven state does not exist
    field(EIVL, "8")
    field(EIST, "FTHRM")
#    field(FLNK, "$(P):MathMode1_RBV")
}

record(mbbo, "$(P):MathMode2") {
    field(DESC, "Math mode 2")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "CONT")
    # Two state does not exist
    field(THVL, "3")
    field(THST, "CTHRM")
    field(FRVL, "4")
    field(FRST, "DB")
    field(FVVL, "5")
    field(FVST, "DBM")
    field(SXVL, "6")
    field(SXST, "FILTER")
    # Seven state does not exist
    field(EIVL, "8")
    field(EIST, "FTHRM")
    field(FLNK, "$(P):MathMode1")
}

record(mbbi, "$(P):MathMode1_RBV") {
    field(DESC, "Math mode")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getMathMode(MATH?, $(P)) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "CONT")
    # Two state does not exist
    field(THVL, "3")
    field(THST, "CTHRM")
    field(FRVL, "4")
    field(FRST, "DB")
    field(FVVL, "5")
    field(FVST, "DBM")
    field(SXVL, "6")
    field(SXST, "FILTER")
    # Seven state does not exist
    field(EIVL, "8")
    field(EIST, "FTHRM")
}

record(mbbi, "$(P):MathMode2_RBV") {
    field(DESC, "Math mode")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "CONT")
    # Two state does not exist
    field(THVL, "3")
    field(THST, "CTHRM")
    field(FRVL, "4")
    field(FRST, "DB")
    field(FVVL, "5")
    field(FVST, "DBM")
    field(SXVL, "6")
    field(SXST, "FILTER")
    # Seven state does not exist
    field(EIVL, "8")
    field(EIST, "FTHRM")
}
#--- End Math mode --------


record(bo, "$(P):KeyboardLock") {
    field(DESC, "Keyboard lock")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setInt(LOCK, $(P)) $(PORT) $(ADDR)")
    field(ZNAM, "OFF")
    field(ONAM, "ON")
#    field(FLNK, "$(P):KeyboardLock_RBV")
}

record(bi, "$(P):KeyboardLock_RBV") {
    field(DESC, "Keyboard lock")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(LOCK?) $(PORT) $(ADDR)")
    field(ZNAM, "OFF")
    field(ONAM, "ON")
}


record(ao, "$(P):Timer") {
    field(DESC, "Timer sample interval")
    field(DTYP, "stream")
    field(LOPR, "0")
    field(HOPR, "6000")
    field(DRVL, "0")
    field(DRVH, "6000")
    field(EGU,  "s")
    field(PREC, "6")
    field(OUT,  "@$(PROTOFILE) setFloat(TIMER, $(P)) $(PORT) $(ADDR)")
#    field(FLNK, "$(P):Timer_RBV")
}

record(ai, "$(P):Timer_RBV") {
    field(DESC, "Timer sample interval")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getFloat(TIMER?) $(PORT) $(ADDR)")
    field(EGU,  "s")
    field(PREC, "6")
}


record(mbbo, "$(P):MemMode") {
    field(DESC, "Mem mode")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setInt(MEM, $(P)) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "LIFO")
    field(TWVL, "2")
    field(TWST, "FIFO")
    field(THVL, "3")
    field(THST, "CONT")
#    field(FLNK, "$(P):MemMode_RBV")
}

record(mbbi, "$(P):MemMode_RBV") {
    field(DESC, "Mem mode")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(MEM?) $(PORT) $(ADDR)")
    field(ZRVL, "0")
    field(ZRST, "OFF")
    field(ONVL, "1")
    field(ONST, "LIFO")
    field(TWVL, "2")
    field(TWST, "FIFO")
    field(THVL, "3")
    field(THST, "CONT")
}


record(mbbo, "$(P):MemFormat") {
    field(DESC, "Mem mode")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setInt(MFORMAT, $(P)) $(PORT) $(ADDR)")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "ASCII")
    field(TWVL, "2")
    field(TWST, "SINT")
    field(THVL, "3")
    field(THST, "DINT")
    field(FRVL, "4")
    field(FRST, "SREAL")
    field(FVVL, "5")
    field(FVST, "DREAL")
#    field(FLNK, "$(P):MemFormat_RBV")
}

record(mbbi, "$(P):MemFormat_RBV") {
    field(DESC, "Mem mode")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(MFORMAT?) $(PORT) $(ADDR)")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "ASCII")
    field(TWVL, "2")
    field(TWST, "SINT")
    field(THVL, "3")
    field(THST, "DINT")
    field(FRVL, "4")
    field(FRST, "SREAL")
    field(FVVL, "5")
    field(FVST, "DREAL")
}


record(mbbo, "$(P):TrigArmMode") {
    field(DESC, "Mem mode")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setInt(TARM, $(P)) $(PORT) $(ADDR)")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "AUTO")
    field(TWVL, "2")
    field(TWST, "EXT")
    field(THVL, "3")
    field(THST, "SGL")
    field(FRVL, "4")
    field(FRST, "HOLD")
    field(FVVL, "5")
    field(FVST, "SYN")
#    field(FLNK, "$(P):TrigArmModeWait")
}

record(seq, "$(P):TrigArmModeWait") {
    field(DESC, "")
    field(DLY1, "5.0")
    field(LNK1, "$(P):TrigArmMode_RBV.PROC")
}

record(mbbi, "$(P):TrigArmMode_RBV") {
    field(DESC, "Mem mode")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(TARM?) $(PORT) $(ADDR)")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "AUTO")
    field(TWVL, "2")
    field(TWST, "EXT")
    field(THVL, "3")
    field(THST, "SGL")
    field(FRVL, "4")
    field(FRST, "HOLD")
    field(FVVL, "5")
    field(FVST, "SYN")
}


#--- Num readings (NRDGS) --------
record(longout, "$(P):NumReadings") {
    field(DESC, "Number of readings")
    field(LOPR, "1")
    field(HOPR, "16777215")
    field(DRVL, "1")
    field(DRVH, "16777215")
    field(FLNK, "$(P):NumReadingsMode")
}

record(longin, "$(P):NumReadings_RBV") {
    field(DESC, "Number of readings")
}

record(mbbo, "$(P):NumReadingsMode") {
    field(DESC, "Number of readings")
    field(DTYP, "stream")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "AUTO")
    field(TWVL, "2")
    field(TWST, "EXTSYN")
    # Three state does not exist
    # Four state does not exist
    field(FVVL, "5")
    field(FVST, "SYN")
    field(SXVL, "6")
    field(SXST, "TIMER")
    field(SVVL, "7")
    field(SVST, "LEVEL")
    field(EIVL, "8")
    field(EIST, "LINE")
    field(OUT,  "@$(PROTOFILE) setNumReadings(NRDGS, $(P)) $(PORT) $(ADDR)")
#    field(FLNK, "$(P):NumReadingsMode_RBV")
}

record(mbbi, "$(P):NumReadingsMode_RBV") {
    field(DESC, "Number of readings")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getNumReadings(NRDGS?, $(P)) $(PORT) $(ADDR)")
    # Zero state does not exist
    field(ONVL, "1")
    field(ONST, "AUTO")
    field(TWVL, "2")
    field(TWST, "EXTSYN")
    # Three state does not exist
    # Four state does not exist
    field(FVVL, "5")
    field(FVST, "SYN")
    field(SXVL, "6")
    field(SXST, "TIMER")
    field(SVVL, "7")
    field(SVST, "LEVEL")
    field(EIVL, "8")
    field(EIST, "LINE")
}
#--- End Num readings (NRDGS) --------

record(bo, "$(P):InputBuffer") {
    field(DESC, "Keyboard lock")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) setInt(INBUF, $(P)) $(PORT) $(ADDR)")
    field(ZNAM, "OFF")
    field(ONAM, "ON")
#    field(FLNK, "$(P):InputBuffer_RBV")
}

record(bi, "$(P):InputBuffer_RBV") {
    field(DESC, "Keyboard lock")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(INBUF?) $(PORT) $(ADDR)")
    field(ZNAM, "OFF")
    field(ONAM, "ON")
}


record(longin, "$(P):ReadingCount_RBV") {
    field(DESC, "Stored reading count")
    field(DTYP, "stream")
    field(INP,  "@$(PROTOFILE) getInt(MCOUNT?) $(PORT) $(ADDR)")
}


record(longout, "$(P):MemRecall") {
    field(DESC, "Recall memory")
    field(DTYP, "stream")
    field(OUT,  "@$(PROTOFILE) getMem($(P):Mem_RBV) $(PORT) $(ADDR)")
}


record(waveform, "$(P):Mem_RBV") {
    field(DESC, "Memory contents")
    field(FTVL, "DOUBLE")
    field(NELM, "1000")
}

record(compress, "$(P):MemAveComp") {
    field(DESC, "Memory contents average")
    field(INP,  "$(P):Mem_RBV")
    field(ALG,  "Average")
    field(FLNK, "$(P):MemAve")
}

record(ai, "$(P):MemAve") {
    field(DESC, "Memory contents average")
    field(INP,  "$(P):MemAveComp")
    field(PREC, "6")
}


#--- High-level sequences -------------------------
record(seq, "$(P):Initialize") {
    field(DESC, "Initialize")
    field(DLY1, "0.0")
    field(LNK1, "$(P):Preset.PROC")
    field(DLY2, "$(DLY1=0.25)")
    field(LNK2, "$(P):LineFreq.PROC")
}

record(seq, "$(P):MeasModeResistance") {
    field(DESC, "Set meas to: OHMS,auto,0.001%")
    field(DLY1, "0.0")
    field(DO1,  "4")
    field(LNK1, "$(P):MeasMode PP")
    field(DLY2, "$(DLY1=0.25)")
    field(DO2,  "-1")
    field(LNK2, "$(P):MeasRange PP")
    field(DLY3, "$(DLY1=0.25)")
    field(DO3,  "0.001")
    field(LNK3, "$(P):MeasRes PP")
    field(DLY4, "$(DLY1=0.25)")
    field(LNK4, "$(P):MeasSet.PROC")
}

record(seq, "$(P):MeasModeDCV") {
    field(DESC, "Set meas to: DCV,10.0,auto")
    field(DLY1, "0.0")
    field(DO1,  "1")
    field(LNK1, "$(P):MeasMode PP")
    field(DLY2, "$(DLY1=0.25)")
    field(DO2,  "10.0")
    field(LNK2, "$(P):MeasRange PP")
    field(DLY3, "$(DLY1=0.25)")
    field(DO3,  "-1")
    field(LNK3, "$(P):MeasRes PP")
    field(DLY4, "$(DLY1=0.25)")
    field(LNK4, "$(P):MeasSet.PROC")
}

record(seq, "$(P):GetVoltage") {
    field(DESC, "Get voltage")
#    field(DLY1, "0.0")
#    field(LNK1, "$(P):Preset.PROC")
    field(DLY2, "$(DLYG=1.0)")
    field(DO2,  "10")
    field(LNK2, "$(P):IntTime PP")
    field(DLY3, "$(DLYG=1.0)")
    field(DO3,  "1")
    field(LNK3, "$(P):MeasMode PP")
    field(DLY4, "$(DLYR=1.0)")
    field(LNK4, "$(P):MeasSet.PROC")
    field(DLY5, "$(DLYG=1.0)")
    field(DO5,  "3")
    field(LNK5, "$(P):TrigData PP")
}

record(seq, "$(P):GetResistance") {
    field(DESC, "Get resistance")
#    field(DLY1, "0.0")
#    field(LNK1, "$(P):Preset.PROC")
    field(DLY2, "$(DLYR=1.0)")
    field(DO2,  "10")
    field(LNK2, "$(P):IntTime PP")
    field(DLY3, "$(DLYR=1.0)")
    field(LNK3, "$(P):MeasModeResistance.PROC")
    field(DLY4, "$(DLYR=1.0)")
    field(DO4,  "3")
    field(LNK4, "$(P):TrigData PP")
}


record(seq, "$(P):SetUpTimedVMeas1") {
    field(DESC, "Set up timed voltage meas.")
    # Preset the HP3458
    # Scott A. says this is not needed
#    field(DLY1, "0.0")
#    field(LNK1, "$(P):Preset.PROC")
    # Turn off autorange
    field(DLY2, "$(DLYSV=0.5)")
    field(DO2,  "0")
    field(LNK2, "$(P):AutorangeMode PP")
    # Turn off the autozero
    field(DLY3, "$(DLYSV=0.5)")
    field(DO3,  "0")
    field(LNK3, "$(P):AutozeroMode PP")
    # Turn off the display
    field(DLY4, "$(DLYSV=0.5)")
    field(DO4,  "0")
    field(LNK4, "$(P):DisplayMode PP")
    # Turn off math functions
    field(DLY5, "$(DLYSV=0.5)")
    field(DO5,  "0")
    field(LNK5, "$(P):MathMode1 PP")
    field(DLY6, "$(DLYSV=0.5)")
    field(DO6,  "0")
    field(LNK6, "$(P):MathMode2 PP")
    # Turn on the keyboard lock
    field(DLY7, "$(DLYSV=0.5)")
    field(DO7,  "1")
    field(LNK7, "$(P):KeyboardLock PP")
    # Set the output format to ASCII
    field(DLY8, "$(DLYSV=0.5)")
    field(DO8,  "1")
    field(LNK8, "$(P):OutFormat PP")
    # Set the reference frequency (power line) 
    field(DLY9, "$(DLYSV=0.5)")
    field(DO9,  "1")
    field(LNK9, "$(P):LineFreq PP")
    # Set trig to HOLD 
    field(DLYA, "$(DLYSV=0.5)")
    field(DOA,  "4")
    field(LNKA, "$(P):TrigMode PP")
    field(FLNK, "$(P):SetUpTimedVMeas2")
}

record(seq, "$(P):SetUpTimedVMeas2") {
    field(DESC, "Set up timed voltage meas.")
    # Set measurement mode
    field(DLY1, "0.5")
    field(LNK1, "$(P):MeasModeDCV.PROC")
    # Set the A/D integration time (in units of NPLC)
    field(DLY2, "1.0")
    field(DO2,  "2")
    field(LNK2, "$(P):IntTime PP")
    # Set the timer
    field(DLY3, "$(DLYSV=0.5)")
    field(DO3,  "3.3380E-2")
    field(LNK3, "$(P):Timer PP")
    # Enable memory and set mode to FIFO
    field(DLY4, "$(DLYSV=0.5)")
    field(DO4,  "2")
    field(LNK4, "$(P):MemMode PP")
    # Clear memory and set format to DINT
    field(DLY5, "$(DLYSV=0.5)")
    field(DO5,  "3")
    field(LNK5, "$(P):MemFormat PP")
    # Set up the trigger 
    field(DLY6, "$(DLYSV=0.5)")
    field(DO6,  "1")
    field(LNK6, "$(P):TrigArmMode PP")
    # Set number of readings
    field(DLY7, "$(DLYSV=0.5)")
    field(DO7,  "719")
    field(LNK7, "$(P):NumReadings PP")
    # Set num readings mode to TIMER
    field(DLY8, "$(DLYSV=0.5)")
    field(DO8,  "6")
    field(LNK8, "$(P):NumReadingsMode PP")
}

record(seq, "$(P):TriggerTimedVMeas") {
    field(DESC, "Trigger timed voltage meas.")
    # Turn on the input buffer
    field(DLY1, "0.0")
    field(DO1,  "1")
    field(LNK1, "$(P):InputBuffer PP")
    # Trigger the readings
    field(DLY2, "1.0")
    field(DO2,  "3")
    field(LNK2, "$(P):TrigMode PP")
    # Check for errors
    field(DLY3, "1.0")
    field(LNK3, "$(P):Error.PROC")
    # Check status
    field(DLY4, "0.5")
    field(LNK4, "$(P):Status.PROC")
}

record(seq, "$(P):CollectTimedVMeas") {
    field(DESC, "Collect timed voltage meas.")
    # Check the number of readings in memory
    field(DLY1, "0.0")
    field(LNK1, "$(P):ReadingCount_RBV.PROC")
    # Put all the readings in the HP3458 output buffer
    field(DLY2, "1.0")
    field(DOL2, "$(P):ReadingCount_RBV")
    field(LNK2, "$(P):MemRecall PP")
    # Check for errors
    field(DLY3, "1.0")
    field(LNK3, "$(P):Error.PROC")
    # Check status
    field(DLY4, "0.5")
    field(LNK4, "$(P):Status.PROC")
}
#--- End High-level sequences -------------------------


#--- Record processing sequences -------------------------
record(seq, "$(P):SeqInit1") {
    field(DLY1, "1.0")
    field(LNK1, "$(P):ID.PROC")
    field(DLY2, "$(DLY1=0.25)")
    field(LNK2, "$(P):Error.PROC")
    field(DLY3, "$(DLY1=0.25)")
    field(LNK3, "$(P):IntTime_RBV.PROC")
    field(DLY4, "$(DLY1=0.25)")
    field(LNK4, "$(P):OutFormat_RBV.PROC")
    field(DLY5, "$(DLY1=0.25)")
    field(LNK5, "$(P):LineFreq_RBV.PROC")
    field(DLY6, "$(DLY1=0.25)")
    field(LNK6, "$(P):TrigMode_RBV.PROC")
    field(DLY7, "$(DLY1=0.25)")
    field(LNK7, "$(P):IntegerScale_RBV.PROC")
    field(DLY8, "$(DLY1=0.25)")
    field(LNK8, "$(P):MeasMode_RBV.PROC")
    field(DLY9, "$(DLY1=0.25)")
    field(LNK9, "$(P):AutorangeMode_RBV.PROC")
    field(DLYA, "$(DLY1=0.25)")
    field(LNKA, "$(P):AutozeroMode_RBV.PROC")
    field(PINI, "1")
    field(FLNK, "$(P):SeqInit2")
}

record(seq, "$(P):SeqInit2") {
    field(DLY1, "$(DLY1=0.25)")
    field(LNK1, "$(P):DisplayMode_RBV.PROC")
    field(DLY2, "$(DLY1=0.25)")
    field(LNK2, "$(P):MathMode1_RBV.PROC")
    field(DLY3, "$(DLY1=0.25)")
    field(LNK3, "$(P):KeyboardLock_RBV.PROC")
    field(DLY4, "$(DLY1=0.25)")
    field(LNK4, "$(P):Timer_RBV.PROC")
    field(DLY5, "$(DLY1=0.25)")
    field(LNK5, "$(P):MemMode_RBV.PROC")
    field(DLY6, "$(DLY1=0.25)")
    field(LNK6, "$(P):MemFormat_RBV.PROC")
    field(DLY7, "$(DLY1=0.25)")
    field(LNK7, "$(P):TrigArmMode_RBV.PROC")
    field(DLY8, "$(DLY1=0.25)")
    field(LNK8, "$(P):NumReadingsMode_RBV.PROC")
    field(DLY9, "$(DLY1=0.25)")
    field(LNK9, "$(P):InputBuffer_RBV.PROC")
    field(DLYA, "$(DLY1=0.25)")
    field(LNKA, "$(P):ReadingCount_RBV.PROC")
    field(FLNK, "$(P):SeqInit3")
}

record(seq, "$(P):SeqInit3") {
    field(DLY1, "$(DLY1=0.25)")
    field(LNK1, "$(P):Status.PROC")
}

record(seq, "$(P):SeqSlow1") {
    field(DLY1, "0.0")
    field(LNK1, "$(P):SeqInit1.PROC")
#    field(SCAN, "10 second")
}

record(seq, "$(P):SeqFast1") {
    field(DLY1, "0.0")
    field(LNK1, "$(P):Error.PROC")
    field(DLY2, "$(DLY1=0.25)")
    field(LNK2, "$(P):ReadingCount_RBV.PROC")
    field(DLY3, "$(DLY1=0.25)")
    field(LNK3, "$(P):Status.PROC")
#    field(SCAN, "5 second")
}
#--- End Record processing sequences -------------------------