[SLAC] [SLAC Pubs and Reports]
SLAC-PUB-7906
Characterization of Low-Energy (6-30 keV) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities
Abstract
Relative to 137Cs gamma rays, the response of MTS-N
(NatLiF:Mg,Ti-0.04 cm thick) increased from 0.4 to 1.4
between x-ray energies of 6 and 30 keV and that of MCP-N
(NatLif:Mg, Cu, P-0.04 cm thick) increased from 0.02 to 1.2
between 6 and 26 keV. A mathematical model for TLD response was used
with measured responses to determine overall TLD efficiency C(k) and
efficiency relative to 137Cs. The effective energy attenuation
coefficient mu(k) and light attenuation coefficient f were also
determined experimentally. The relative efficiency of MTS-N increased
from 0.66 to 1.30 between 6 and 30 keV and that of MCP-N increased
from 0.37 to 0.99 between 6 and 26 keV.
Full Text
PDF
Compressed PostScript
More Information
Full bibliographic data for this document, including its complete author list,
is (or soon will be) available from SLAC's
SPIRES-HEP
Database.
Please report problems with this file to
posting@slac.stanford.edu.
The SLAC preprint inventory is provided by the
SLAC Technical Publications Department.
Page generated 04 Apr 2001 @ 15:08 PDT by htmlme.pl