SLAC-PUB-7828
Studies of Beam Position Monitor Stability

Abstract

We present the results from two studies of the time stability between the mechanical center of a beam position monitor and its electrical/electronic center. In the first study, a group of 93 BPM processors was calibrated via Test Pulse Generator once per hour in order to measure the contribution of the readout electronics to offset drifts. In the second study, a triplet of stripline BPMs in the Final Focus Test Beam, separated only by drift spaces, was read out every 6 minutes during 1 week of beam operation. In both cases offset stability was observed to be on the order of microns over time spans ranging from hours to days, although during the beam study much worse performance was also observed. Implications for the beam position monitor system of future linear collider systems are discussed.

Full Text

PDF

Compressed PostScript

Notes

The PDF version of this document was distilled from a PostScript file most likely created from LaTeX source which used Type 3 fonts. The resultant PDF may be unreadable on the screen at the default viewing magnification (fonts will appear bitmapped, jagged), though the type will be legible at higher magnifications.

Despite the cosmetic problems resulting from the use Type 3 fonts in the original source file, such PDF files print without problems; likewise, the onscreen text is searchable and selectable.

More Information

Full bibliographic data for this document, including its complete author list, is (or soon will be) available from SLAC's SPIRES-HEP Database.

Please report problems with this file to posting@slac.stanford.edu. The SLAC preprint inventory is provided by the SLAC Technical Publications Department.
Page generated 04 Apr 2001 @ 15:08 PDT by htmlme.pl