Turner's 1-6 PAU calibration procedure
To calibrate 1-6 PAU:
- Do a Klys ENLD FTP (LI01 KLYS PANEL, select 1-6, DIAGNOSTIC
PANEL, KLYS EGAIN FTP)
- Print it. You want the cursor value.
- Max out the 1-6 AMPL's (Second KLYS INDEX, 1-6 PULSED AMPLs,
611, 613, 614 to their upper limits)
- Select AMPL 611.
- Do a saturation FTP for 1-6 (LI01 KLYS PANEL)
- Raise the drive of 1-6 so the cursor falls just below the knee
of the saturation curve (ENTER DRIVE % to just to the
left of where the curve starts to level off).
- From the CRR application Panel, go to Select ENERGY VERNER.
- SETUP VERNER KLYS --> enter "LI01 61" for the sampled variable
name and "11" for the beamcode for PAU.
- START AUTO ACQ. (takes about 10 minutes)
- ANALYZ KLYS DATA. If it looks reasonable (i.e. the E_NOLOAD
goes up to 280 MeV and the curve fits nicely), ACCEPT
ENERGY POLY.
- Knob the 1-6 AMPL's down, checking them against ENLD FTP's
for 1-6, until the cursor value is close to what it
was originally.