BaBar EMC detector grounding and shielding
This document contains a description of the final calorimeter grounding and shielding after installation in IR2. It is a bringing together of the information on many circuit diagrams.
Original Grounding and shielding plan set out by John Dowdell
Final Barrel Calorimeter grounding and shielding
General description
The grounding system for the barrel calorimeter is a "star" topology. The centre of the star is the cylinder-minicrate-bulkhead assembly.
There are one main path for the ground, from the IOB power supply grounds to the preamp through the IOB, ADB, bulkhead and FOB. There is one main shielding path from the copper cooling block to the pre-amp box via the IOB and ADB cans, bulkhead and
FOB. The ground and shield are connected together through a 10 ohm resistor on the bulkhead board. However a second resistor on the IOB can be used to make this connection if it proves beneficial.
The cylinder is connected to the nominal detector ground. Whilst another leg of the "star" goes to the TRB and EMB
Around the entire system is the outer shield which is presently separate from the inner shielding scheme.
All the power supplies are floating being referenced to ground through the route described in the diagram below.
Diagram
PostScript Version
/BFROOT/www/Detector/DAQ/Calorimeter/WriteUps/EMCground.ps
Index to diagram
- The three power lines, +5V, -5V and +10V are floating at their
supplies in the electronics house. Their returns are joined together on the
IOB.
- Power (digital) ground is passed from the IOB to the ADBs on
many pins of the two ERNI connectors.
- The IOB can and each ADB can are connected through pins 1 and
2 of the 50 way ERNI connector.
- The analogue and digital grounds are connected together under
pin 44 of each of the ADCs on the ADB.
- The analogue ground is passed onto the bulkhead board, twice per
pre-amp (A and B) down the middle row of the AMP 250 pin connector (24 pins).
The bias grounds enter the bulkhead on a four pin connector. The bulkhead
passes the ground to the ADB on pins 50a and 50e of the AMP connector.
The ADB then ties the bias ground to analogue ground.
- The ADB can is connected to the bulhead using the AMP connector
shield. The AMP shield is conneted to the upper and lower metal layer of the
bulkhead board using many plated through holes.
- The upper and lower metal layers of the bulkhead make contact with
the Minicrate when the crate is screwed together around the bulkhead board.
(Since the minicrates were anodysed this coating had to be removed along the
inner edge where good electrical contact had to be made to the bulkhead board)
- The analogue ground passes straight through the bulkhead board
on the pins of the AMP connector, (refer to E above)
- The upper and lower metal layers of the bulkhead are connected to
5 pins (1a,1b,1c,1d,1e) at one end of the 250 pin AMP connector. This ground
is then fanned out to pins 1 and 22 of the pre amp cable connectors through
a 10 Ohm resistor. Which is connected to the cable shield drain wire.
- The Minicrate makes electrical contact with the barrel along the
inner surface of its fixing flange (not anodysed).
- At the bottom of the forward end of the barrel a cable is bolted
to the cylinder and tied to the nominal detector ground.
- Each TRB can has a connection to the cylinder through a cable
attached at on end to the TRB power cable clamp and at the other end to a
minicrate fixing bolt. The TRB cans are mounted on the solenoid but make no
electrical contact with it, being insulated by a sheet of thermally
conducting foam.
- The TRB power cables are clamped to the TRB can. In a small cable
clamp.
- The three power lines, +5V, -5V and +10V are floating at their
supplies in the electronics house. Their returns are joined together on the
TRB. The electrical ground
- The pre-amp cable shield is connected to the pre-amp box with
a small hinged cable clamp. The last 2cm of shield has no insulating covering
for this purpose.
- The copper cooling blocks are bolted to the IOB cans. The heat path
is through the cylinder so in order to create a grounding loop a thin sheet
of electrically insulating material is placed between the copper blocks and
the minicrate flange to which the copper blocks are bolted.
- The EMB derives its ground from the TRB through th 14 way ribbon
cable.
Final Endcap Calorimeter grounding and shielding
Diagram
Index to diagram
Pete Sanders
Last modified: Thursday October 8 1998
|