Summary of Prototype II Running
Extracted from proto2data.html 23 July 2001
Information last updated 25 Jun 1998 by Masahiro Morii
<masahiro@slac.stanford.edu>
- Phase 1
- Colorado
- Phase 2
- SLAC + single-hit TDC
- Phase 3
- SLAC + multi-hit TDC
- Phase 4
- FEA
- Phase 5
- FEA + new DCAC
See also Proto-II
Data Structure, Proto-II
Cosmic Run Procedure and Proto-II TDC
Threshold.
Phase 1 - Runs 381-394
Phase 1 data were recorded using the Colorado preamplifier boards. The discriminator threshold was fixed at 300mV.
Run Start date/time Events Tape.seq HV Comments
===========================================================================
381 31-Jan-97/11:35 98 QX0142.1 --- Calibration
382 31-Jan-97/13:30 870 QX0142.2 1919V Cosmic, HV -5%
383 31-Jan-97/14:11 430 QX0142.3 1919V (trigger problem?)
384 31-Jan-97/14:37 57 QX0142.4 1919V (trigger problem?)
385 31-Jan-97/15:04 1490 QX0142.5 1919V
386 31-Jan-97/17:23 25051 QX0142.6 1919V
387 01-Feb-97/11:16 9772 QX0142.7 1919V
388 01-Feb-97/18:07 16 QX0142.8 1919V (trigger problem?)
389 01-Feb-97/18:09 24171 QX0142.9 1919V
390 02-Feb-97/11:01 8739 QX0142.10 1919V
391 02-Feb-97/17:16 22653 QX0142.11 1919V
392 03-Feb-97/19:53 19167 QX0142.12 2020V Cosmic, HV "nominal"
393 04-Feb-97/19:34 22994 QX0142.13 2020V
394 05-Feb-97/19:05 20820 QX0142.14 2020V
===========================================================================
Runs 381 onward were recorded after "fixing" the TDCs by distributing common stop signal at 950ns after each trigger. All TDCs should be working. Event flags are incorrect; ignore them.
Run 381 is a calibration run (but flagged as cosmic data in the file).
Runs with smaller sizes were stopped because of suspected trigger problems (e.g. trigger rates were too high). No clear problems were found, but they may well be discarded.
HV setting for runs 382-391 (called "-5%") was: 1919V on anodes, 865V and 333V on guard wires. This voltage, 5% lower than the nominal voltage calculated using DCSIM, was found to be close to the nominal gas gain of 5x10^4.
HV setting for runs 392-394 (called "nominal") was 2020V on anodes, 910V and 350V on guard. This is the nominal voltage calculated by DCSIM. The gas gain was measured to be about 1-1.2x10^5.
Phase 2 - Runs 682-706
Phase 2 data were recorded using
- New preamplifier board. Colorado -> SLAC.
This means different threshold and different t0.
- DAQ software mods. Calibration and slow data integrated.
Run Start date/time Events Tape.seq HV Comments
===========================================================================
682 26-Mar-97/19:43 20003 QX0142.15 2020V
684 27-Mar-97/18:21 20187 QX0142.16 2020V slow data included
685 28-Mar-97/18:59 21500 QX0142.17 2020V
686 29-Mar-97/10:38 42064 QX0142.18 2020V
689 30-Mar-97/17:54 21965 QX0142.19 2020V
690 31-Mar-97/19:20 20479 QX0142.20 2020V
691 01-Apr-97/18:23 21429 QX0142.21 2020V
700 02-Apr-97/18:30 21267 QX0142.22 2020V HV -> slow data
702 03-Apr-97/18:31 20657 QX0142.23 2020V
703 04-Apr-97/18:19 24747 QX0142.24 2020V
705 05-Apr-97/12:14 33398 QX0142.25 2020V
706 06-Apr-97/13:36 27418 QX0142.26 2020V
===========================================================================
HV setting is now 2020V/910V/350V by default. The gas gain is re-estimated to be 1.8x10^5 on average, although there is significant difference between layers. Discriminator threshold was set to 260mV. This corresponds to about 1.5 electrons at the above
HV setting.
Calibration scheme has changed so that only 1 out of 4 channels per preamp chip is calibrated in each event. This makes 4 (channels) x 10 (delay) x 10 = 400 events for the calibration at the beginning of each run.
Slow control data (temperature, humidity, etc.) have been incorporated into the data stream starting from run 684. The SlowData structure changed (unfortunately):
- Runs 684-691: 16 variables. There were three variables for HV, which were not read and filled with zeros.
- Runs 700-706: 17 variables. There were four variables for HV, which were actually filled. The additional one (cathode 4) is for the outer guard (910V) wires.
The latter structure will (hopefully) be kept during the next data taking period.
Data files for runs 689 onward are smaller by a factor 1/3 because channels with no TDC hits were suppressed. (We should've done this long ago...)
Data taking was stopped after run 706 in order to move the chamber from B288 to B273. Total number of triggers = 295,114.
Phase 3 - Runs 847-959
Phase 3 data were recorded using the SLAC preamplifier board and LeCroy multi-hit TDCs. See Changes between run 706 and run847.
Run Start date/time Events Tape.seq HV Comments
===========================================================================
847 30-May-97/12:05 3772 QX0142.27 2020V short waveform
849 30-May-97/19:40 12422 QX0142.28 2020V
850 31-May-97/16:02 14105 QX0142.29 2020V FADC problem
851 01-Jun-97/15:02 11424 QX0142.30 2020V FADC problem
852 02-Jun-97/09:22 15283 QX0142.31 2020V FADC problem
853 03-Jun-97/10:04 15311 QX0142.32 2020V FADC problem
865 04-Jun-97/20:02 9356 QX0142.33 2020V FADC problem
899 05-Jun-97/18:47 9789 QX0142.34 2020V
920 07-Jun-97/21:55 8623 QX0142.35 2020V
921 08-Jun-97/11:31 19207 QX0142.36 2020V
922 09-Jun-97/18:26 23751 QX0142.37 2020V unusually long run
924 11-Jun-97/13:40 9695 QX0142.38 2020V Trig counters moved backward
925 12-Jun-97/18:45 15677 QX0142.39 2020V
926 14-Jun-97/19:46 14814 QX0142.40 2020V
927 16-Jun-97/18:11 6389 QX0142.41 2020V
928 17-Jun-97/14:45 7451 QX0142.42 2020V
930 18-Jun-97/15:48 16013 QX0142.43 2020V lower PMT drawing large current
931 23-Jun-97/09:30 7269 QX0142.44 2020V
932 25-Jun-97/14:15 14761 QX0142.45 2020V Trig counters moved to middle
935 27-Jun-97/10:50 1403 QX0142.46 2020V
949 27-Jun-97/17:15 20628 QX0142.47 2020V pedestal problem
952 30-Jun-97/09:23 1618 QX0142.49 2020V
959 30-Jun-97/18:05 5485 QX0142.48 2020V
===========================================================================
HV (2020V/910V/350V) and discriminator threshold (260mV) remain the same as the previous period. Calibration scheme is also kept the same, although the multi-hit TDC used in
this period has different range and sensitivity from the single-hit TDCs of the previous period.
FADC waveforms were recorded in this period. The nominal length of the recorded waveform is 200ns (10 x 20ns) before and 1000ns (50 x 20ns) after each `signal'. There was a bug in Run 847 which made the waveform shorter, 200ns before and 800ns after
each signal.
Runs 850-865 had one FADC module with pedestal problem. The module covers four wires with layer/wire = 14/34, 14/36, 16/34 and 16/36. The waveform from these channels were junk.
After run 922 the iron stack was moved 23 inches toward the electronics. The electronics for layers 5-12 (superlayers 2 and 3) shifted by two cells.
Run 931 contains the last data taken with the iron stack at the end of the prototype. After this run, the stack was moved back to the centre of the chamber and the electronics was appropriately shifted.
Phase 4 - Runs 1175-1182
Phase 4 data were recorded using the preproduction Front-End Assemblies. See Proto-II Cosmic Run Phase 4 for more details.
Run Start date/time Events Tape.seq HV Comments
===========================================================================
1175 29-Aug-97/15:36 523 QX0142.50 2020V Vth=300mV, TDC trig data
1176 29-Aug-97/16:51 393 QX0142.51 2020V Vth=300mV, FADC=3 trig data
1179 29-Aug-97/18:28 4069 QX0142.52 2020V Vth=300mV
1180 30-Aug-97/11:50 5475 QX0142.53 2020V Vth=260mV
1181 31-Aug-97/09:42 8343 QX0142.54 2020V Vth=260mV
1182 01-Sep-97/18:41 3781 QX0142.55 2020V Vth=260mV
===========================================================================
There are 320 calibration events at the beginning of each run. The sequence of these events are:
- For 4 channel mask patterns (0001, 0010, 0100, 1000) {
- For 4 strobe timings (17, 34, 51, 68ns) {
- Generate 10 pulses
}
- For 4 amplitudes (100, 200, 300, 400mV) {
- Generate 10 pulses
}
}
Run 1175 and 1176 are short runs for testing the trigger data stream. The discriminator threshold was set to 300mV. The trigger was generated using TDC hits (run 1175) or FADC delta > 3 (run 1176).
Run 1179-1182 are cosmic data for normal resolution studies etc. The discriminator threshold was set to 300mV for run 1179, and to 260mV for run 1180-1182. The electronics was reasonably quiet (only a few noisy channels at 260mV), which was achieved by
a dirty workaround using copper tapes et al.
Phase 5 - Runs 1236-1355
Phase 5 data were recorded using the preproduction DCACs and modified FEAs. See Proto-II Cosmic Run Phase 5 for more details.
Run Start date/time Events Tape.seq HV Comments
===========================================================================
1236 14-Nov-97/19:40 9061 QX0142.56 2020V Vth=220mV
1237 15-Nov-97/14:37 20510 QX0142.57 2020V Events after 9898 empty
1238 QX0142.58 2020V No reconstructed tracks
1239 QX0142.59 2020V DGAIN high, FEA cables swapped
1260 QX0142.60 2020V DGAIN low, HV trip: isobutane empty
1302 QX0142.61 2020V DGAIN low, BUG: outer elefants killed
1328 25-Nov-97/17:00 11405 QX0142.62 2020V DGAIN high
1330 26-Nov-97/10:02 17428 QX0142.63 2020V DGAIN low
1332 28-Nov-97/13:53 14805 QX0142.64 1960V DGAIN low, gas gain 1/2
===========================================================================
Earlier part of this phase was not very successful due to numerous hardware/software glitches. You should look at the later part, starting from run 1338, for useful data.
Run 1236 and 1237 were taken with two ADBs, one in each FEA. The trigger counters were near the rear end plate to compensate for the limited acceptance of the electronics. The preamplifiers (preproduction) were used with the default gain and full
digital bandwidth. The 220mV threshold resulted in a noise hit rate of about 0.3 hit/wire. The data has only 32 calibration events at the beginning of run. This was done similar to the 320 events calibration described above, but with only 1 pulse at each
point.
There was a problem during run 1237 after event 9898, which made all subsequent events empty. The run has therefore only 9898 useful events.
The trigger counters were moved back to the center of the chamber before run 1238. This run seems to have had hardware problem; there is no tracks.
Run 1239 was taken with the digital gain raised by a factor of 2. There was a cabling problem which severely limited the geometrical acceptance.
Run 1260 had a gas problem: we ran out of isobutane. The gas gain went high until the HV tripped.
Run 1302 had a problem with software which disabled the outer ADBs. This resulted in almost no geometrical coverage.
Run 1332 was taken with lower HV: anode 1960V, zipper 340 and guard 880V. This should lower the gas gain by a factor of 2.
There was a significant hardware modification between run 1332 and run 1338. The Elefants were found to be measuring the trailing edge of the TDC inputs due to misdefinition of the signal polarity. The signals were flipped so that the leading
edge timing would be measured. This problem was affecting all the data since Phase 4.
Run Start date/time Events Tape.seq HV Comments
===========================================================================
1338 01-Dec-97/21:58 9757 QX0142.65 2020V DGAIN low
1347 02-Dec-97/16:13 11969 QX0142.66 2020V DGAIN high
1348 03-Dec-97/11:59 7322 QX0142.67 1960V DGAIN high
1349 03-Dec-97/23:00 9349 QX0142.68 1960V DGAIN high
1350 04-Dec-97/19:03 9688 QX0142.69 1900V DGAIN high, FADC gain=20
1351 05-Dec-97/17:35 11633 QX0142.70 1900V DGAIN high, FADC gain=20, CALDAC high
1353 06-Dec-97/16:32 12206 QX0142.71 2020V DGAIN high, FADC gain=20
1354 07-Dec-97/19:06 10148 QX0142.72 1900V DGAIN high, FADC gain=20
1355 23-Dec-97/14:39 3325 QX0142.73 1960V trigger recorded
1374 10-Feb-98/18:04 10784 QX0142.74 1960V New and old ADBs
===========================================================================
Runs 1338 onwards, after the hardware fix, turned out to be very good. Two FEAs (both fully populated with preproduction chips) were used to cover rather large geometrical acceptance. The trigger counter was in the middle. DGAIN of the preamp
was switched between low and high, the latter being the expected working point for the real chamber. Three HV points were tried: 2020V as a reference (same as many old runs), 1960V and 1900V for more realistic working points. The absolute gas gain is yet
to be determined, but the gain ratios are believed to be about G(1960V)/G(2020)V = 0.60 and G(1900V)/G(2020)V = 0.35.
For most runs, the analog gain of the Elefant was set to 5 (nominal). We tried to change it to 10 for runs 1350 and 1351, where the low HV setting would have make it difficult to measure the waveform. Due to a software problem, we ended up changing the
analog gain to 20. The FADC was always set to linear mode (V_top = 4.0V, V_mid = 2.5V, V_bot = 1.0V).
In run 1351, CALDAC was set to the maximum value (4.095V). This was found to reduce the noise by ~20mV for some yet unknown reason. With DGAIN set to high, we could run at threshold = 260mV compared with 300mV in run 1350. It is not yet clear if this
really improved the S/N ratio; it might just have reduced the digital gain.
Run 1353 was meant to be a repeat of run 1347, but I forgot to set the FADC gain back to 5. It should not matter as far as the spatial resolution is concerned.
Run 1355 is a short run for testing trigger bits. The condition was: HV=1960V, dgain high, FADC gain (still) 20, threshold 300mV. CALDAC was normal (100mV).
Run 1374 is a test run for the new (3rd generation) ADBs. Two FEA boxes equipped with old (lefti box) and new (right box) ADBs were used to take the data at HV = 1960V and DGAIN = low. The digital gain was not completely uniform: One of the old ADBs
had a diode on the DGAIN line making it open when the DGAIN is set low; other ADBs (1 old and 2 new) had 33k resistors to the CMOS driver making the gain a little lower. The threshold was 200mV, CALDAC 4V, FADC gain 10 (I believe).
Run Start date/time Events Tape.seq HV Comments
===========================================================================
1390 28-Feb-98/17:38 11160 QX0142.75 1960V \
1391 01-Mar-98/11:43 13433 QX0143.1 2020V \
1392 02-Mar-98/17:48 13098 QX0143.2 1900V \ Junk
1393 03-Mar-98/18:53 13703 QX0143.3 1960V / Threshold set to 500mV
1394 04-Mar-98/21:42 15971 QX0143.4 2020V /
1395 05-Mar-98/18:45 12730 QX0143.5 1900V /
1397 06-Mar-98/19:23 15309 QX0143.6 1960V \ 3rd generation FEA
1398 07-Mar-98/17:58 13565 QX0143.7 2020V > Pedestal calibration
1399 08-Mar-98/14:15 13582 QX0143.8 1900V / Bilinear FADC
1410 04-Apr-98/12:58 18870 QX0143.9 1960V FADC gain 10
1411 05-Apr-98/14:49 14414 QX0143.10 1960V FADC gain 5
===========================================================================
Runs 1390 onwards used two pre-production FEAs fully loaded with 3rd generation ADBs. The running conditions were chosen to be as close as possible to what we will use on the real chamber. DGAIN = low. threshold = 200mV. FADC was set to bi-linear mode,
i.e., Vtop = 4.0V, Vmid = 1.75V, Vbot = 1.0V. Vref was set to 1.1V to bring the pedestal around 5 FADC count.
In order to facilitate the determination of the FADC pedestal in each event, the timing of the Level-1 Accept was delayed by 16 sysclks (=267ns) compared to the previous runs. This should move the waveforms by 4 FADC samples, leaving more room before
the signal. A side effect of this change is that the measured hit time is now shifted by 267ns and the T0 must be retuned.
We also started to do pedestal calibration at the beginning of each run. This consists of 110 events with no signal (readout sparsification was disabled, of course). During the pedestal calibration, Vref was scanned within ±50mV of the
nominal value in 11 steps, changing the pedestal value by ±2-3 counts. Averaging the pedestal values over these 11 Vref values should wash out the quantization error and give us a better measurement of the true pedestal.
Unfortunately, a bug was introduced in the start-of-run procedure which set the threshold to 500mV for the first 6 runs of this period. Runs 1390-1395 are therefore junk. The bug was fixed before Run 1397.
From Run 1397 onwards, the TDC calibration was done with the "small-signal" calibration mode, i.e. SEL33X = 1, and CALDAC set to 100mV. The FADC calibration was done in the "large-signal" mode with CALDAC scanning from 50mV to 200mV in 50mV steps.
Data taking resumed from Run 1410 after a month of break. The condition was kept same: 3rd generation ADB with preproduction DCACs, "low" DGAIN, threshold at 200mV. HV was kept at the nominal 1960V.
Run Start date/time Events Tape.seq HV Comments
===========================================================================
1420 11-Apr-98/22:41 10676 QX0143.11 2020V \ New DCAC, thresh = 200mV
1421 12-Apr-98/21:57 10803 QX0143.12 2020V / Junk runs
1431 13-Apr-98/20:28 11584 QX0143.13 2020V thresh 200mV, gas problem?
1433 14-Apr-98/18:46 11436 QX0143.14 1960V thresh 180mV, 2 final ADBs
1435 16-Apr-98/00:18 10506 QX0143.15 1960V thresh 180mV, 3 final ADBs
1437 17-Apr-98/19:10 24295 QX0143.16 2020V \ 4 final ADBs
1438 18-Apr-98/16:36 13624 QX0143.17 1960V > thresh 180mV
1439 19-Apr-98/11:14 20221 QX0143.18 1960V /
===========================================================================
Runs 1420 onwards used Production DCAC. First two runs, 1420 and 1421, were taken before the chip's peculiarities were understood, and are therefore junk. HV was set to 2020V and the threshold was 200mV.
Significant progress had been made in understanding the final DCAC before Run 1431. The stability of the chips had been greatly improved by setting the common-mode voltage of the threshold inputs at 2.75V instead of 2.4V.
The first run after this breakthrough, Run 1431, turned out to be a disappointement. We suspect that the gas mixture was not right at the beginning (gas flow was set very low), and the calibration changed during the run. There was also a problem with
one ADB in which the FET bias voltage was wrong and the FADC waveforms were lost. HV was set to 2020V and the threshold was 200mV.
Runs 1433 and 1435 were taken at HV = 1960V and the threshold of 180mV. The data quality turned out to be good. Up to this point, we had a mixture of final and pre-final ADBs. The threshold was limited by the pre-final ADBs which showed somewhat poorer
signal-to-noise ratios.
Run 1437 onwards were taken with all final ADBs. At this point, all the components of the FEAs were final. The digital-gain control resistors (33kohm on DGAIN pin of DCAC) were removed, resulting in about 20% increase in the digital gain. The threshold
remained at 180mV. HV was set to 2020V for Run 1437, 1960V for Runs 1438 and 1439.
Run Start date/time Events Tape.seq HV Comments
===========================================================================
1510 03-Jun-98/11:16 27508 QX0413.19 2020V He-CO2-C4H10, Ele gain 5
1614 17-Jun-98/19:22 22433 QX0413.20 2020V He-CO2-C4H10, Ele gain 10
1640 19-Jun-98/17:11 29059 QX0413.21 2100V He-CO2-C4H10, Ele gain 10
===========================================================================
Run 1510 and 1614 were taken with a non-flammable gas mixture, He-CO2-C4H10. HV was set to 2020V, discriminator threshold 180mV. Elefant analog gain was 5 for Run 1510 and 10 for Run 1614. Run 1640 was taken with HV=2100V.
Run Start date/time Events Tape.seq HV Comments
===========================================================================
1792 05-Aug-98/09:07 759 1960V Gas test run
===========================================================================
Run 1792 was taken for checking the quality of a new isobutane bottle. The
``LAB'' DAQ system with one ROM was used to record this data. The
discriminator threshold was 250mV to suppress the excess noise seen only
with ROM+DIOM on Proto-II. The gas mixture is close to 80:20, but may not be
very accurate. The FADC gain was set to 10.
|